Página 1 - Organic Carbon Analyzer

FG7005002ANATEL PAT700 On-line TotalOrganic Carbon AnalyzerUSER MANUALApril 2010, Edition 4© HACH Company, 2007, 2008, 2010. All rights reserved. Prin

Página 2

Page 2 Before You Begin – PAT700 Total Organic Carbon AnalyzerOperator Manual AnatelA touch-screen interface provides access to all analyzer function

Página 3 - Table of Contents

Page 92 Bottle Mode – PAT700 Total Organic Carbon AnalyzerOperator Manual Anatel• C of A value text box allows you to enter the certified analysis va

Página 4 - Chapter 5 Setup

PAT700 Total Organic Carbon Analyzer – Bottle Mode Page 93Anatel Operator ManualFigure 6-9 Test running dialog box• After verifying the presence of t

Página 5

Page 94 Bottle Mode – PAT700 Total Organic Carbon AnalyzerOperator Manual AnatelNOTEImmediately upon completion of the test the bottles are backflush

Página 6

PAT700 Total Organic Carbon Analyzer – Bottle Mode Page 95Anatel Operator ManualFigure 6-12 Unload/replace bottles dialog box• Unlock icon energizes

Página 7

Page 96 Bottle Mode – PAT700 Total Organic Carbon AnalyzerOperator Manual Anatel6.6 Conductivity calibrationCAUTIONPotential puncture hazard! Contact

Página 8

PAT700 Total Organic Carbon Analyzer – Bottle Mode Page 97Anatel Operator Manual11) Install the standards bottle following the instructions as shown

Página 9 - Chapter 1 Before You Begin

Page 98 Bottle Mode – PAT700 Total Organic Carbon AnalyzerOperator Manual AnatelFigure 6-15 Verification resistor installed on printer/calibration po

Página 10 - 1.3 RFID technology

PAT700 Total Organic Carbon Analyzer – Bottle Mode Page 99Anatel Operator Manual• Cancel icon takes you to a dialog box that notifies you that the te

Página 11 - 1.5 Safety

Page 100 Bottle Mode – PAT700 Total Organic Carbon AnalyzerOperator Manual Anatel6.7 TOC validationCAUTIONPotential puncture hazard! Contact with the

Página 12 - 1.7 FCC conformance

PAT700 Total Organic Carbon Analyzer – Bottle Mode Page 101Anatel Operator ManualFigure 6-18 Load bottles dialog box, TOC validation• Bottle position

Página 13 - 1.8 EU directives

PAT700 Total Organic Carbon Analyzer – Before You Begin Page 3Anatel Operator ManualThe reader/writer retrieves the data stored on the RFID tags loca

Página 14 - 1.9 Precautionary labels

Page 102 Bottle Mode – PAT700 Total Organic Carbon AnalyzerOperator Manual AnatelNOTEImmediately upon completion of the test the bottles are backflus

Página 15 - 1.12 Definitions of terms

PAT700 Total Organic Carbon Analyzer – Bottle Mode Page 103Anatel Operator ManualFigure 6-19 Load bottles dialog box, system suitability test• Bottle

Página 16 - 1.13 Customer service

Page 104 Bottle Mode – PAT700 Total Organic Carbon AnalyzerOperator Manual AnatelWhere:SR = Suitability response of the analyzer defined as (rss – rw

Página 17 - Chapter 2 Installation

PAT700 Total Organic Carbon Analyzer – Bottle Mode Page 105Anatel Operator Manual• Bottle 3 = empty• Bottle 4 = emptyFigure 6-20 Load bottles dialog

Página 18 - 2.2 Orientation and mounting

Page 106 Bottle Mode – PAT700 Total Organic Carbon AnalyzerOperator Manual Anatel• Reps text box allows you to specify the number of repetitions to p

Página 19

PAT700 Total Organic Carbon Analyzer – Bottle Mode Page 107Anatel Operator ManualFigure 6-23 Grab sample summary dialog boxExport icon opens the expo

Página 20 - Mounting to a wall

Page 108 Bottle Mode – PAT700 Total Organic Carbon AnalyzerOperator Manual Anatel6.10 Excursion modeNOTEExcursion mode requires the use of an RFID ta

Página 21

PAT700 Total Organic Carbon Analyzer – Bottle Mode Page 109Anatel Operator ManualExcursion with validationCAUTIONPotential puncture hazard! Contact w

Página 22

Page 110 Bottle Mode – PAT700 Total Organic Carbon AnalyzerOperator Manual AnatelFigure 6-27 Excursion mode setup dialog box• TOC limit text box allo

Página 23 - Isolation valve

PAT700 Total Organic Carbon Analyzer – Bottle Mode Page 111Anatel Operator Manual• Following the excursion and validation, a validation results dialo

Página 24 - Water inlet and outlet

Page 4 Before You Begin – PAT700 Total Organic Carbon AnalyzerOperator Manual Anatel• Follow all procedures in Return procedures, page 171, before sh

Página 25

Page 112 Bottle Mode – PAT700 Total Organic Carbon AnalyzerOperator Manual AnatelFigure 6-30 Excursion mode without validation dialog box• Bottle pos

Página 26 - 2.4 Setting sample flow rate

PAT700 Total Organic Carbon Analyzer – Bottle Mode Page 113Anatel Operator Manual• Run now icon immediately draws a water sample from the online sour

Página 27

Page 114 Bottle Mode – PAT700 Total Organic Carbon AnalyzerOperator Manual AnatelFigure 6-33 Checking bottles dialog boxThe analyzer will start the b

Página 28 - Power supply wiring

PAT700 Total Organic Carbon Analyzer – Alarms Page 115Anatel Operator ManualChapter 7 Alarms7.1 Alarm indicationThe analyzer has two basic operationa

Página 29 - I/O wiring

Page 116 Alarms – PAT700 Total Organic Carbon AnalyzerOperator Manual Anatel7.2 Alarm acknowledgement1) If an alarm occurs, at the home screen, touch

Página 30 - Discrete inputs

PAT700 Total Organic Carbon Analyzer – Alarms Page 117Anatel Operator Manual7.3 Alarm detailsTouching one of the alarms in the list produces the alar

Página 31

Page 118 Alarms – PAT700 Total Organic Carbon AnalyzerOperator Manual AnatelError alarmsErrors occur any time one of the non-critical subsystems fail

Página 32 - Discrete outputs

PAT700 Total Organic Carbon Analyzer – Alarms Page 119Anatel Operator ManualCritical error alarmsA critical error occurs any time the analyzer cannot

Página 33

Page 120 Alarms – PAT700 Total Organic Carbon AnalyzerOperator Manual AnatelSC104 NotEnoughFluidinBottle1 • Not enough fluid in bottle 1 to complete

Página 34

PAT700 Total Organic Carbon Analyzer – Alarms Page 121Anatel Operator ManualSC141 UnableToAcknowledgeAlarms Analyzer can not acknowledge alarm.SC142

Página 35

PAT700 Total Organic Carbon Analyzer – Before You Begin Page 5Anatel Operator ManualThis equipment has been tested and found to comply with the limit

Página 36

Page 122 Alarms – PAT700 Total Organic Carbon AnalyzerOperator Manual AnatelTOC alarmsA TOC or conductivity alarm occurs any time the measured TOC le

Página 37

PAT700 Total Organic Carbon Analyzer – Diagnostics Page 123Anatel Operator ManualChapter 8 Diagnostics8.1 Accessing the diagnostics dialog box• At th

Página 38 - 2.7 Serial ports

Page 124 Diagnostics – PAT700 Total Organic Carbon AnalyzerOperator Manual AnatelLamp test icon turns on each lamp for 180 seconds and tests the UV o

Página 39 - Conventions

PAT700 Total Organic Carbon Analyzer – Diagnostics Page 125Anatel Operator ManualPAT700 lamp replacementTo replace a lamp, follow these steps:1) Touc

Página 40 - Data read commands

Page 126 Diagnostics – PAT700 Total Organic Carbon AnalyzerOperator Manual AnatelFigure 8-4 Diagnostics dialog box, test tabRS-232 testThe RS-232 tes

Página 41 - Data logger commands

PAT700 Total Organic Carbon Analyzer – Diagnostics Page 127Anatel Operator ManualDigital I/O testThe digital I/O test icon allows you to test digital

Página 42

Page 128 Diagnostics – PAT700 Total Organic Carbon AnalyzerOperator Manual AnatelFigure 8-7 4-20 mA output test dialog boxOutput 1, output 2, and out

Página 43 - Parameter setup command

PAT700 Total Organic Carbon Analyzer – Diagnostics Page 129Anatel Operator ManualFigure 8-8 Plumbing test dialog boxFlow path selection drop down lis

Página 44

Page 130 Diagnostics – PAT700 Total Organic Carbon AnalyzerOperator Manual AnatelFigure 8-9 Select bottles to load dialog box, plumbing testCAUTIONPo

Página 45

PAT700 Total Organic Carbon Analyzer – Diagnostics Page 131Anatel Operator ManualFigure 8-10 Load bottles dialog box, plumbing test• Bottle positions

Página 46 - Diagnostic commands

Page 6 Before You Begin – PAT700 Total Organic Carbon AnalyzerOperator Manual Anatel1.9 Precautionary labelsRead all labels and tags attached to the

Página 47

Page 132 Diagnostics – PAT700 Total Organic Carbon AnalyzerOperator Manual Anatel6) If a test is chosen immediately, the analyzer asks if you wish to

Página 48 - Auto logoff

PAT700 Total Organic Carbon Analyzer – Diagnostics Page 133Anatel Operator ManualFigure 8-13 Pump calibration dialog boxTest time text box indicates

Página 49 - Sliding toolbar

Page 134 Diagnostics – PAT700 Total Organic Carbon AnalyzerOperator Manual AnatelFigure 8-14 Fan test dialog boxRFID testThe RFID test icon allows yo

Página 50

PAT700 Total Organic Carbon Analyzer – Diagnostics Page 135Anatel Operator Manual8.5 Test tabScreen calibration icon allows the user to calibrate the

Página 51 - Current tab

Page 136 Diagnostics – PAT700 Total Organic Carbon AnalyzerOperator Manual Anatel

Página 52

PAT700 Total Organic Carbon Analyzer – Data Review Page 137Anatel Operator ManualChapter 9 Data Review9.1 Accessing the data review dialog box1) At t

Página 53 - Graph tab

Page 138 Data Review – PAT700 Total Organic Carbon AnalyzerOperator Manual Anatel9.3 Export dataData are exported in formatted text. The export funct

Página 54

PAT700 Total Organic Carbon Analyzer – Data Review Page 139Anatel Operator Manual9.4 Filter dataData filtering allows you to select the type and rang

Página 55

Page 140 Data Review – PAT700 Total Organic Carbon AnalyzerOperator Manual Anatel

Página 56 - 4.3 Online TOC mode

PAT700 Total Organic Carbon Analyzer – Signing On and Off Page 141Anatel Operator ManualChapter 10 Signing On and OffIf security is enabled, to acces

Página 57 - 4.4 Conductivity mode

PAT700 Total Organic Carbon Analyzer – Before You Begin Page 7Anatel Operator Manual1.10 Symbols and marksThe PAT700 Total Organic Carbon Analyzer co

Página 58 - 4.6 Offline mode

Page 142 Signing On and Off – PAT700 Total Organic Carbon AnalyzerOperator Manual Anatel10.2 PasswordFigure 10-1 Password dialog boxUser ID. Enter yo

Página 59 - 4.8 Self clean mode

PAT700 Total Organic Carbon Analyzer – Modbus Protocol Page 143Anatel Operator ManualChapter 11 Modbus Protocol11.1 IntroductionModbusThis chapter pr

Página 60

Page 144 Modbus Protocol – PAT700 Total Organic Carbon AnalyzerOperator Manual Anatel11.2 Modbus/TCP DriverModbus ProtocolA typical Modbus/TCP frame

Página 61

PAT700 Total Organic Carbon Analyzer – Modbus Protocol Page 145Anatel Operator ManualRequestResponse*N = quantity of registersErrorAny other Function

Página 62 - General tab

Page 146 Modbus Protocol – PAT700 Total Organic Carbon AnalyzerOperator Manual AnatelData ModelModbus bases its data model on a series of tables that

Página 63 - Idle tab

PAT700 Total Organic Carbon Analyzer – Modbus Protocol Page 147Anatel Operator ManualData EncodingModbus uses a 'big-endian' representation

Página 64 - Digital control tab

Page 148 Modbus Protocol – PAT700 Total Organic Carbon AnalyzerOperator Manual AnatelDoubleA Double float value is 64-bits within the Anatel PAT700 T

Página 65 - Stabilization tab

PAT700 Total Organic Carbon Analyzer – Modbus Protocol Page 149Anatel Operator ManualDateThe Date is transmitted similar to a 32-bit Word with the MO

Página 66 - P5 Curve

Page 150 Modbus Protocol – PAT700 Total Organic Carbon AnalyzerOperator Manual AnatelData DictionaryThe following tables detail the Modbus addresses

Página 67 - 5.4 System

PAT700 Total Organic Carbon Analyzer – Modbus Protocol Page 151Anatel Operator ManualAddress 15-24 (Model Number) (Format: Strings)Get the model numb

Página 68 - Display tab

Page 8 Before You Begin – PAT700 Total Organic Carbon AnalyzerOperator Manual AnatelParts per billion (ppb). A term used to report trace chemical ana

Página 69 - Network tab

Page 152 Modbus Protocol – PAT700 Total Organic Carbon AnalyzerOperator Manual AnatelConfiguration BlockThe Configuration Data block comprises parame

Página 70 - Backlight tab

PAT700 Total Organic Carbon Analyzer – Modbus Protocol Page 153Anatel Operator ManualAddress 100-101 (Instrument IP Address) (Format: IP Address)Get

Página 71 - 5.5 Alarms

Page 154 Modbus Protocol – PAT700 Total Organic Carbon AnalyzerOperator Manual AnatelAddress 102-103 (Subnet Mask) (Format: IP Address)Get the instru

Página 72

PAT700 Total Organic Carbon Analyzer – Modbus Protocol Page 155Anatel Operator ManualAddress 118 (Idle Mode) (Format: 16 Bit Word)Get / Set the instr

Página 73 - 5.6 Analogs

Page 156 Modbus Protocol – PAT700 Total Organic Carbon AnalyzerOperator Manual AnatelAddress 136-139 (Cell calibration offset) (Format: Double Float)

Página 74 - 5.7 Printer

PAT700 Total Organic Carbon Analyzer – Modbus Protocol Page 157Anatel Operator ManualAddress 166 (TOC Alarm) (Format: 16 Bit Word)Get / Set the instr

Página 75

Page 158 Modbus Protocol – PAT700 Total Organic Carbon AnalyzerOperator Manual AnatelAddress 192-195 (Analog Cond. Full scale) (Format: Double Float)

Página 76

PAT700 Total Organic Carbon Analyzer – Modbus Protocol Page 159Anatel Operator ManualLast TOC Analysis Parameters BlockThe Last TOC Analysis Paramete

Página 77 - 5.8 Security

Page 160 Modbus Protocol – PAT700 Total Organic Carbon AnalyzerOperator Manual AnatelUser Security BlockThe User Security block comprises of paramete

Página 78

PAT700 Total Organic Carbon Analyzer – Modbus Protocol Page 161Anatel Operator ManualAddress 427 (Data logging) (Format: 16 Bit Word)Get / Set the in

Página 79 - Settings tab

PAT700 Total Organic Carbon Analyzer – Installation Page 9Anatel Operator ManualChapter 2 InstallationDANGERTasks involving installation of this equi

Página 80 - Users tab

Page 162 Modbus Protocol – PAT700 Total Organic Carbon AnalyzerOperator Manual Anatel

Página 81

PAT700 Total Organic Carbon Analyzer – Maintenance and Troubleshooting Page 163Anatel Operator ManualChapter 12 Maintenance and TroubleshootingDANGER

Página 82 - Audit trail tab

Page 164 Maintenance and Troubleshooting – PAT700 Total Organic Carbon AnalyzerOperator Manual AnatelFailed lamp indicationsWhen code 128 is produced

Página 83

PAT700 Total Organic Carbon Analyzer – Maintenance and Troubleshooting Page 165Anatel Operator ManualWARNINGUV Lamp related shock, eye/skin, ozone ga

Página 84 - TOC calibration tab

Page 166 Maintenance and Troubleshooting – PAT700 Total Organic Carbon AnalyzerOperator Manual Anatel8) Grasp the lamp module on the right and left s

Página 85 - Conductivity calibration tab

PAT700 Total Organic Carbon Analyzer – Maintenance and Troubleshooting Page 167Anatel Operator Manual9) Press and hold the lamp retaining clip to rel

Página 86 - System suitability tab

Page 168 Maintenance and Troubleshooting – PAT700 Total Organic Carbon AnalyzerOperator Manual Anatel12) Reconnect the lamp power connectors by caref

Página 87 - Grab sample tab

PAT700 Total Organic Carbon Analyzer – Maintenance and Troubleshooting Page 169Anatel Operator Manual12.2 Water filter replacementCAUTIONBurn and pre

Página 88 - 5.10 Password

Page 170 Maintenance and Troubleshooting – PAT700 Total Organic Carbon AnalyzerOperator Manual Anatel12.3 TroubleshootingThe PAT700 incorporates exte

Página 89 - 5.11 Factory

PAT700 Total Organic Carbon Analyzer – Maintenance and Troubleshooting Page 171Anatel Operator Manual• The instrument’s serial number.• Sample resist

Página 90 - Thermistor tab

Page 10 Installation – PAT700 Total Organic Carbon AnalyzerOperator Manual Anatel2.2 Orientation and mountingNOTEFailure to mount the analyzer in an

Página 91 - Algorithm tab

Page 172 Maintenance and Troubleshooting – PAT700 Total Organic Carbon AnalyzerOperator Manual AnatelNOTEFailure to purge the analyzer of all liquid,

Página 92 - OxFiles tab

PAT700 Total Organic Carbon Analyzer – Maintenance and Troubleshooting Page 173Anatel Operator ManualTo purge the PAT700 of entrained water:1) Remove

Página 93 - Chapter 6 Bottle Mode

Page 174 Maintenance and Troubleshooting – PAT700 Total Organic Carbon AnalyzerOperator Manual Anatel

Página 94 - 6.3 Bottle mode dialog box

PAT700 Total Organic Carbon Analyzer – Specifications Page 175Anatel Operator Manual Appendix A: SpecificationsPAT700 specifications are subject to c

Página 95 - 6.4 Run standards

Page 176 Specifications – PAT700 Total Organic Carbon AnalyzerOperator Manual AnatelPhysical specificationsUV lamps Two, with UV Detect technologyInt

Página 96 - TOC calibration setup

PAT700 Total Organic Carbon Analyzer – Specifications Page 177Anatel Operator ManualInput/output specificationsAnalog output specificationsAnalog out

Página 97 - TOC calibration procedure

Page 178 Specifications – PAT700 Total Organic Carbon AnalyzerOperator Manual Anatel

Página 98

PAT700 Total Organic Carbon Analyzer – CE Conformance Page 179Anatel Operator Manual Appendix B: CE Conformance

Página 100 - Operator Manual Anatel

PAT700 Total Organic Carbon Analyzer – Theory of Operation Page 181Anatel Operator Manual Appendix C: Theory of OperationThe Anatel PAT700 on-line TO

Página 101 - Anatel Operator Manual

PAT700 Total Organic Carbon Analyzer – Installation Page 11Anatel Operator ManualFigure 2-1 PAT700 mounting dimensions[597]23.5[228.6]9[548.6]21.6[17

Página 102

Page 182 Theory of Operation – PAT700 Total Organic Carbon AnalyzerOperator Manual Anatelto a radio antenna mounted on a substrate. The RFID tag is a

Página 103

PAT700 Total Organic Carbon Analyzer – PAT700 Default Settings Page 183Anatel Operator Manual Appendix D: PAT700 Default SettingsTable D-1 Default s

Página 104 - 6.6 Conductivity calibration

Page 184 PAT700 Default Settings – PAT700 Total Organic Carbon AnalyzerOperator Manual AnatelAnalog Outputs Setup Default SettingTOC zero BlankTOC fu

Página 105

PAT700 Total Organic Carbon Analyzer – PAT700 Default Settings Page 185Anatel Operator ManualBottles Change Settings Default SettingTOC calibrationMa

Página 106 - Verification resistor

Page 186 PAT700 Default Settings – PAT700 Total Organic Carbon AnalyzerOperator Manual Anatel

Página 108 - 6.7 TOC validation

Page 12 Installation – PAT700 Total Organic Carbon AnalyzerOperator Manual AnatelFigure 2-2 PAT700 analyzer orientationMounting general consideration

Página 109

PAT700 Total Organic Carbon Analyzer – Installation Page 13Anatel Operator ManualFigure 2-3 PAT700 wall mountMounting to dual instrument polesFollow

Página 110 - 6.8 System suitability test

Page 14 Installation – PAT700 Total Organic Carbon AnalyzerOperator Manual Anatel• Use four 5/16-inch U-bolts, two for each 2-inch pipe, and eight ma

Página 111

PAT700 Total Organic Carbon Analyzer – Installation Page 15Anatel Operator Manual2.3 Plumbing connectionsCAUTIONPotential eye and/or burn hazard! The

Página 112 - 6.9 Grab sample

Page 16 Installation – PAT700 Total Organic Carbon AnalyzerOperator Manual AnatelFigure 2-5 PAT700 connection to isolation valveWater inlet and outle

Página 113

PAT700 Total Organic Carbon Analyzer – Installation Page 17Anatel Operator ManualFigure 2-6 PAT700 water inlet and outlet connectionsTo connect plumb

Página 114

Page 18 Installation – PAT700 Total Organic Carbon AnalyzerOperator Manual Anatel2.4 Setting sample flow rateThe PAT700 is designed to operate with a

Página 115

PAT700 Total Organic Carbon Analyzer – Installation Page 19Anatel Operator Manual2.5 Wiring connections for PAT700 with conduit openingsRefer to this

Página 116 - 6.10 Excursion mode

Page 20 Installation – PAT700 Total Organic Carbon AnalyzerOperator Manual AnatelPower supply wiringWARNINGPotential property damage or electrical sh

Página 117 - Excursion with validation

PAT700 Total Organic Carbon Analyzer – Installation Page 21Anatel Operator ManualI/O wiringThe analyzer has three 4-20 mA outputs, two discrete input

Página 118

PAT700 Total Organic Carbon Analyzer – Table of Contents Page iAnatel Operator ManualTable of ContentsChapter 1 Before You Begin1.1 About this manual

Página 119 - Excursion without validation

Page 22 Installation – PAT700 Total Organic Carbon AnalyzerOperator Manual AnatelDiscrete inputsThe analyzer has two discrete inputs, referred to as

Página 120

PAT700 Total Organic Carbon Analyzer – Installation Page 23Anatel Operator ManualFigure 2-11 Digital input wiring external power - IsolatedFigure 2-1

Página 121 - 6.11 Bottle flush mode

Page 24 Installation – PAT700 Total Organic Carbon AnalyzerOperator Manual AnatelDiscrete outputsThe analyzer has four discrete outputs. Wiring conne

Página 122

PAT700 Total Organic Carbon Analyzer – Installation Page 25Anatel Operator Manual2.6 Wiring connections for PAT700 with quick-connect fittingsRefer t

Página 123 - Chapter 7 Alarms

Page 26 Installation – PAT700 Total Organic Carbon AnalyzerOperator Manual AnatelTable 2-6 lists specifications for the power supply.I/O wiringThe an

Página 124 - 7.2 Alarm acknowledgement

PAT700 Total Organic Carbon Analyzer – Installation Page 27Anatel Operator ManualDiscrete inputsThe analyzer has two discrete inputs, referred to as

Página 125 - 7.4 Alarm codes

Page 28 Installation – PAT700 Total Organic Carbon AnalyzerOperator Manual AnatelFigure 2-14 Digital input wiring external power - IsolatedFigure 2-1

Página 126 - Error alarms

PAT700 Total Organic Carbon Analyzer – Installation Page 29Anatel Operator ManualDiscrete outputsThe analyzer has four discrete outputs, as listed in

Página 127 - Warning alarms

Page 30 Installation – PAT700 Total Organic Carbon AnalyzerOperator Manual Anatel2.7 Serial portsThe analyzer has the following serial ports:• A fema

Página 128

PAT700 Total Organic Carbon Analyzer – Installation Page 31Anatel Operator Manual2.8 RS-232 commandsCommands consist of 2-character, ASCII text mnemo

Página 129

Page ii Table of Contents – PAT700 Total Organic Carbon AnalyzerOperator Manual AnatelDiagnostic commands . . . . . . . . . . . . . . . . . . . . .

Página 130 - Audible alarms

Page 32 Installation – PAT700 Total Organic Carbon AnalyzerOperator Manual AnatelMode setup commandsData read commandsThe data normally are displayed

Página 131 - Chapter 8 Diagnostics

PAT700 Total Organic Carbon Analyzer – Installation Page 33Anatel Operator ManualWhere:Data logger commandsExample:Read command: “RE”Response:Alarm t

Página 132

Page 34 Installation – PAT700 Total Organic Carbon AnalyzerOperator Manual AnatelExample:Read command:“LP” or “LP 2007 07 25 12 13 2007 07 25 16 00”R

Página 133 - 8.4 Tests tab

PAT700 Total Organic Carbon Analyzer – Installation Page 35Anatel Operator ManualParameter setup commandExample:Read command: “SY”Response:“725200720

Página 134 - RS-232 test

Page 36 Installation – PAT700 Total Organic Carbon AnalyzerOperator Manual AnatelExample:Read command: “HR”Read format:“HH:mm:SS HH:mm:SS u u u u s u

Página 135 - 4-20 mA output test

PAT700 Total Organic Carbon Analyzer – Installation Page 37Anatel Operator ManualExample:Read command: “PC”Response:Example:Read command: “PS”Respons

Página 136 - Plumbing test

Page 38 Installation – PAT700 Total Organic Carbon AnalyzerOperator Manual AnatelExample:Read command: “PV”Response:Diagnostic commandsExample:Read c

Página 137

PAT700 Total Organic Carbon Analyzer – Startup Page 39Anatel Operator ManualChapter 3 Startup3.1 Startup sequenceThe analyzer goes through the follow

Página 138

Page 40 Startup – PAT700 Total Organic Carbon AnalyzerOperator Manual AnatelNOTETo ensure accurate and representative results after installation or l

Página 139

PAT700 Total Organic Carbon Analyzer – Startup Page 41Anatel Operator Manual3.4 Home screenThe home screen provides an overview of the analysis resul

Página 140 - Pump test

PAT700 Total Organic Carbon Analyzer – Table of Contents Page iiiAnatel Operator Manual5.6 Analogs . . . . . . . . . . . . . . . . . . . . . . . . .

Página 141 - Fan test

Page 42 Startup – PAT700 Total Organic Carbon AnalyzerOperator Manual AnatelFigure 3-2 Home screen with sliding toolbar openRun mode icon allows you

Página 142 - RFID test

PAT700 Total Organic Carbon Analyzer – Startup Page 43Anatel Operator Manual• If no user is signed on, the sign on icon is present. Touching the icon

Página 143 - 8.6 Calibration dates tab

Page 44 Startup – PAT700 Total Organic Carbon AnalyzerOperator Manual Anatel• Temperature is the last temperature reading in “Temperature: X.X °C” fo

Página 144

PAT700 Total Organic Carbon Analyzer – Startup Page 45Anatel Operator ManualLog view tabThe log view tab displays all analysis readings or events as

Página 145 - Chapter 9 Data Review

Page 46 Startup – PAT700 Total Organic Carbon AnalyzerOperator Manual Anatel

Página 146 - 9.3 Export data

PAT700 Total Organic Carbon Analyzer – Run Modes Page 47Anatel Operator ManualChapter 4 Run Modes4.1 The analysis cycleAn analysis cycle is comprised

Página 147 - 9.4 Filter data

Page 48 Run Modes – PAT700 Total Organic Carbon AnalyzerOperator Manual Anatel4.2 Accessing run modesTo access run modes, follow these steps:1) At th

Página 148

PAT700 Total Organic Carbon Analyzer – Run Modes Page 49Anatel Operator ManualSampling with pressureCAUTIONPotential burst hazard! Do not exceed the

Página 149 - Chapter 10 Signing On and Off

Page 50 Run Modes – PAT700 Total Organic Carbon AnalyzerOperator Manual AnatelIn conductivity mode, the analyzer starts by verifying that the UV lamp

Página 150 - 10.4 Backdoor password

PAT700 Total Organic Carbon Analyzer – Run Modes Page 51Anatel Operator Manual4.7 Manual TOC SampleA TOC analysis can be performed manually by touchi

Página 151 - Chapter 11 Modbus Protocol

Page iv Table of Contents – PAT700 Total Organic Carbon AnalyzerOperator Manual Anatel6.10 Excursion mode . . . . . . . . . . . . . . . . . . . . .

Página 152 - Modbus Protocol

Page 52 Run Modes – PAT700 Total Organic Carbon AnalyzerOperator Manual AnatelIn manual self clean mode, the analyzer starts by setting the elapsed t

Página 153 - TCP/IP Interface

PAT700 Total Organic Carbon Analyzer – Setup Page 53Anatel Operator ManualChapter 5 Setup5.1 Accessing the setup dialog box1) At the home screen, tou

Página 154 - Data Model

Page 54 Setup – PAT700 Total Organic Carbon AnalyzerOperator Manual Anatel5.3 TOCThe TOC icon allows you to modify the settings for the TOC operating

Página 155 - Data Encoding

PAT700 Total Organic Carbon Analyzer – Setup Page 55Anatel Operator ManualAverage last results checkbox averages the last X TOC analyses. The average

Página 156 - IP Address

Page 56 Setup – PAT700 Total Organic Carbon AnalyzerOperator Manual AnatelDigital control tabFigure 5-4 TOC dialog box, digital control tabControl wi

Página 157

PAT700 Total Organic Carbon Analyzer – Setup Page 57Anatel Operator ManualStabilization tabFigure 5-5 TOC dialog box, stabilization tabWhen the analy

Página 158 - Data Dictionary

Page 58 Setup – PAT700 Total Organic Carbon AnalyzerOperator Manual AnatelP5 CurveFigure 5-6 P-5 Curve dialog boxThe user selectable P5 warning alarm

Página 159

PAT700 Total Organic Carbon Analyzer – Setup Page 59Anatel Operator Manual5.4 SystemThe system icon allows you to modify system settings for the anal

Página 160 - Configuration Block

Page 60 Setup – PAT700 Total Organic Carbon AnalyzerOperator Manual AnatelDisplay tabFigure 5-8 System settings dialog box, display tabSet time icon

Página 161

PAT700 Total Organic Carbon Analyzer – Setup Page 61Anatel Operator ManualSounds tabFigure 5-9 System settings dialog box, sounds tabAlarm, warning,

Página 162

PAT700 Total Organic Carbon Analyzer – Table of Contents Page vAnatel Operator Manual10.2 Password . . . . . . . . . . . . . . . . . . . . . . . . .

Página 163

Page 62 Setup – PAT700 Total Organic Carbon AnalyzerOperator Manual AnatelStatic subnet mask text box sets the subnet mask for the analyzer. You may

Página 164 - (Format: Double Float)

PAT700 Total Organic Carbon Analyzer – Setup Page 63Anatel Operator Manual5.5 AlarmsThe alarms icon allows you to enable or disable TOC and conductiv

Página 165

Page 64 Setup – PAT700 Total Organic Carbon AnalyzerOperator Manual AnatelThe actual (uncompensated) conductivity limits are in accordance with the s

Página 166

PAT700 Total Organic Carbon Analyzer – Setup Page 65Anatel Operator Manual5.6 AnalogsThe analogs icon allows you to configure the analog outputs for

Página 167

Page 66 Setup – PAT700 Total Organic Carbon AnalyzerOperator Manual Anatel5.7 PrinterThe printer icon allows you to configure the printer settings fo

Página 168 - User Security Block

PAT700 Total Organic Carbon Analyzer – Setup Page 67Anatel Operator ManualTo change the printer settings:1) Turn the printer power switch OFF.2) Rest

Página 169

Page 68 Setup – PAT700 Total Organic Carbon AnalyzerOperator Manual AnatelTOC tabFigure 5-14 Printer dialog box, TOC tabDisabled, continuous, and per

Página 170

PAT700 Total Organic Carbon Analyzer – Setup Page 69Anatel Operator ManualConductivity tabFigure 5-15 Printer dialog box, conductivity tabDisabled, t

Página 171 - UV Detect technology

Page 70 Setup – PAT700 Total Organic Carbon AnalyzerOperator Manual AnatelGeneral tabFigure 5-16 Security settings dialog box, general tabEnable secu

Página 172 - Lamp replacement procedure

PAT700 Total Organic Carbon Analyzer – Setup Page 71Anatel Operator ManualFilename text box contains the filename for the exported data file, which i

Página 173

Page vi Table of Contents – PAT700 Total Organic Carbon AnalyzerOperator Manual Anatel

Página 174

Page 72 Setup – PAT700 Total Organic Carbon AnalyzerOperator Manual AnatelUsers tabThe users tab allows you to add, edit, or delete users of the anal

Página 175

PAT700 Total Organic Carbon Analyzer – Setup Page 73Anatel Operator ManualPrint security settings icon prints the user settings for all users except

Página 176 - Self clean mode

Page 74 Setup – PAT700 Total Organic Carbon AnalyzerOperator Manual AnatelAudit trail tabOnly administrator and factory users can access the audit tr

Página 177 - 12.2 Water filter replacement

PAT700 Total Organic Carbon Analyzer – Setup Page 75Anatel Operator ManualFilter icon opens the filter data dialog box. See Figure 5-23.Return icon r

Página 178 - Technical support

Page 76 Setup – PAT700 Total Organic Carbon AnalyzerOperator Manual AnatelFrom/to date and time range text box. Only audit trail entries that fall be

Página 179 - Return procedures

PAT700 Total Organic Carbon Analyzer – Setup Page 77Anatel Operator ManualR2 limit text box allows you to specify the acceptable limit for the determ

Página 180 - Draining the analyzer

Page 78 Setup – PAT700 Total Organic Carbon AnalyzerOperator Manual AnatelTOC validation tabFigure 5-26 Bottle test setup dialog box, TOC validation

Página 181

PAT700 Total Organic Carbon Analyzer – Setup Page 79Anatel Operator Manualthe analyzer stays at the bottle mode dialog box. All valves are closed, th

Página 182

Page 80 Setup – PAT700 Total Organic Carbon AnalyzerOperator Manual Anatel5.10 PasswordThe password icon allows you to change your password. For more

Página 183 - Appendix A: Specifications

PAT700 Total Organic Carbon Analyzer – Setup Page 81Anatel Operator Manual5.11 FactoryThe factory icon enables factory personnel to enter information

Página 184 - Physical specifications

PAT700 Total Organic Carbon Analyzer – Before You Begin Page 1Anatel Operator ManualChapter 1 Before You Begin1.1 About this manualThis instruction m

Página 185

Page 82 Setup – PAT700 Total Organic Carbon AnalyzerOperator Manual AnatelCell tabFigure 5-31 Factory settings dialog box, cell tabGain text box (rea

Página 186

PAT700 Total Organic Carbon Analyzer – Setup Page 83Anatel Operator ManualUser slopes tabFigure 5-33 Factory settings dialog box, user slopes tabTOC

Página 187 - Appendix B: CE Conformance

Page 84 Setup – PAT700 Total Organic Carbon AnalyzerOperator Manual AnatelOxFiles tabFigure 5-35 Factory settings dialog box, OxFiles tabThe OxFiles

Página 188

PAT700 Total Organic Carbon Analyzer – Bottle Mode Page 85Anatel Operator ManualChapter 6 Bottle Mode6.1 Calibration and validationBottle mode allows

Página 189

Page 86 Bottle Mode – PAT700 Total Organic Carbon AnalyzerOperator Manual AnatelConductivity calibration protocolPerform a conductivity calibration a

Página 190

PAT700 Total Organic Carbon Analyzer – Bottle Mode Page 87Anatel Operator ManualRun standards button takes you to the run standards dialog box. See R

Página 191

Page 88 Bottle Mode – PAT700 Total Organic Carbon AnalyzerOperator Manual Anatel6.5 TOC calibrationTOC calibration ensures the accuracy of the TOC va

Página 192

PAT700 Total Organic Carbon Analyzer – Bottle Mode Page 89Anatel Operator ManualFigure 6-4 Custom setup dialog boxBottle selection check boxes enable

Página 193

Page 90 Bottle Mode – PAT700 Total Organic Carbon AnalyzerOperator Manual Anatel5) To use default settings, touch the use default settings radio butt

Página 194

PAT700 Total Organic Carbon Analyzer – Bottle Mode Page 91Anatel Operator ManualFigure 6-6 Review test setup dialog box• Test setup text lists the bo